Customization: | Available |
---|---|
Type: | IVD Instrument |
Principle: | Magnetic Beads Method |
Still deciding? Get samples of US$ 8000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
ultrassay™ Ex-96 automatic nucleic acid DNA/RNA extraction system is the latest system launched by our company which been used to automatically extract and purify DNA/RNA, protein, and cells. The magnetic beads and samples are absorbed, transferred and released by the magnetic rod and the magnetic sleeve, Completing automated extraction and purification operations. The operation is automated, fast, and simple. By using a special kit or a 96 deep-well plate, 1 to 96 samples can be operated at the same time. With different kinds of magnetic bead nucleic acid extraction reagents, it can extract
specimens from the animal and plant tissues, blood and body fluids.
Advantage:
Smart Circular Magnetic Binding
Magnetic beads recovery rate > 99%. Greatly reduce the amount of residual magnetic beads in the system
Multi-mix model
3 kinds of amplitude, 6 gears of frequency adjustable, adjustable according to different samples and extraction kits to improve extraction efficiency
Anti-contamination control
Added with 11-level HEPA filter and UV lamp to effectively prevent cross-contamination
Independent heating block
8 independent deep-well heating blocks that can reduce mutual interference between reactions.
Removable magnetic rod
Easy to maintain and replace. Prolong the life of the machine
Technical Specification
Model | ultrassay™ 96 |
---|---|
Principle | Magnetic beads method |
Throughout | 1-96 (192, 24, 48 optional) |
Processing Volume | 20-1000μL (24 optional 50ul-5000ul-10000ul) |
Plate Type | 96 Deep well plate (24 Deep well plate optional) |
Magnetic Rods | 96 (192, 24, 48 optional) |
Collection Efficiency | >99% |
CV | <5% |
Heating | Lysis and elution (room temperature to 125ºC) |
Mixing Model | Slow, medium, fast, 1-6 gears programmable |
Reagent | Open system, suitable for all kinds of reagent |
Extraction Time | 5-45 minutes (depends on project) |
Display | 10″ color graphic display with touch screen, external computer not required |
Protocols | Programmable and storable, quantity up to millions |
Extension Port | 4 USB, ethernet, mouse, keyboard and barcode scanners supported |
Disinfection | UV lamp |
Data Storage | Built-in SSD |
Exhaust | Fan exhaust (Optional side exhaust, rear exhaust, HEPA), Programmable |
Working Noise | 53dBA |
Working Temperature | 10-40ºC |
Working Humidity | 10-90% |